Prosthetic Sensory Feedback Lower Extremity
James H. McElhaney, Ph.D *
Maret G. Maxwell, Ph.D. *
David W. Vaughn, C.P.O. *
Grace Horton, R.P.T. *
Linda Bright, R.N. *
Description
Collection
Title:
Creator:
James H. McElhaney, Ph.D *
Maret G. Maxwell, Ph.D. *
David W. Vaughn, C.P.O. *
Grace Horton, R.P.T. *
Linda Bright, R.N. *
Page Number(s):
Year:
Volume:
Issue:
Text:
Prosthetic Sensory Feedback Lower Extremity
Frank W. Clippinger, M.D. *
James H. McElhaney, Ph.D *
Maret G. Maxwell, Ph.D. *
David W. Vaughn, C.P.O. *
Grace Horton, R.P.T. *
Linda Bright, R.N. *
This is a progress report of a Duke University research project involving sensory feedback from lower extremity amputation prostheses.
It has been assumed for many years that replacement of sensory function in prosthetic limbs was a nearly impossible task. Developments in electronics have made possible small amplifier systems and usable transducers, but the basic difficulty remains that of getting the signals into the central nervous system in a fashion that is interpretable, comfortable, consistent, and convenient.
The problem has not been ignored and the obvious routes-auditory signal, electrical stimulation of intact skin, mechanical stimulation, and developments leading to solving the skin barrier with compatible percutaneous materials have been explored.
From 1969 to 1975, this laboratory developed the mechanism to produce sensation from upper limb prosthetic terminal devices. This system was built around the concept of proportional peripheral nerve stimulation by means of a surgically implanted, induction coupled radio receiver-pulse generator, driven by an external amplifier and transmitter that relayed frequency modulated signals, controlled by a strain gauge transducer in the terminal device.
The conclusions from this study were:
-
The system is feasible and signals can be interpreted with reliability relative to the stimulating activity.
-
The brain interprets the signal as coming from the normal peripheral distribution of the nerve stimulated.
-
Signal threshold and nerve excitability does not deteriorate with time, at least in this application.
-
The implanted device is reliable, and durable, there having been no implant failures in twelve years.
In 1975, a grant was received from the National Cancer Institute to apply this technique to the lower limb amputee. This study is to determine whether sensory feedback, in addition to that provided normally from the stump-socket interface and terminal knee impact, useful or advantageous.
To date, 21 patients have been fitted with a lower extremity sensory feedback system, including below knee, above knee, and hip disarticulation amputees. The majority of these have been cancer patients.
The new amputee from malignancy presents a special problem. It is difficult to subject a person recently amputated for cancer to another surgical procedure to insert a stimulator implant. In addition, the amputation is followed by months of chemotherapy during which time wound healing is compromised and the patient does not feel well. Emotional factors must be considered also.
For this reason, it was necessary to develop a noninvasive system as well as the implanted nerve stimulator. After a brief unsuccessful trial with a skin vibrator, the auditory route was selected.
The electronic systems of both the implanted and auditory devices are similar. The system consists of a set of strain gauges which measure anteroposterior and mediolateral bending moments incorporated into the below knee segment of the prosthesis utilizing an endoskeletal unit developed by the Department of Bioengineering at Duke, hybridized with Ottobock endoskeletal prosthetic components.
In addition to the strain gauges, a pressure activated piezo-electric crystal is imbedded in the heel of a SACH foot. This is activated on heel strike.
When the weight is balanced in mid stance or when the prosthesis is unloaded, as with the patient sitting, there is no signal produced by any of the transducers. The system is designed to provide proportional feedback as soon as weight is biased in any direction.
Figure
For the implant system, the signal to the nerve is frequency-modulated with the frequency of stimulus increasing from 0 to 90 Hertz proportionate to the load. With frequencies greater than 90 Hertz, a decrease in signal or complete loss of signal has been experienced routinely. Voltage is adjusted to a level that is comfortable for the patient. Threshold in these patients has varied between .5 and .9 volts.
The implanted receiver is identical to that used in the upper limb project except that four electrodes are placed around the sciatic nerve in the buttock rather than the two that were used for the median nerve in the upper limb project. The receiver is placed subcutaneously in the lower abdominal wall and the antenna is taped to the overlying skin. Only two electrodes are stimulated and the pair which produces the best response is selected. Electrode orientation is important and this is a compromise. The alternative would be to do the surgery with the patient awake which has obvious disadvantages.
In all patients, an interpretable signal was produced although the mental imaging, which was 90 percent correct in the upper limb, has been haphazard in the lower. No patient has reported that the stimulus or the mental image produced was uncomfortable, unpleasant, or confusing, however.
The auditory system uses the same external transducer unit, but the signal is fed to a hearing aid earpiece placed behind the ear without blocking the external auditory canal.
In that the end result of any sensory feedback is a subjective response, it is difficult to determine its effect in scientific terms.
A gait laboratory has been developed to analyze walking with and without the sensory feedback system. This provides computer-assisted analysis of force plate and segmental accelerometer data. This facet of the study has just started and at the moment, insufficient data analysis is available to be meaningful.
It is felt, however, that the subjective individual patient response will actually be more helpful in the long run. This is "quality of life" response and is voiced as statements like: "I can walk out in the driveway at night without worrying", "I feel better about going downstairs", "I can play basketball better with it turned on", "I can control the accelerator on my car far better".
Not all the subjects have found the system useful. Table I outlines the patients who have had the sensory feedback systems and their outcome. Most of those who have abandoned it, however, have had the auditory unit.
Table I
-
Sensory feedback systems in lower extremity amputees appear to have advantages. How much better the amputees are is still under investigation and whether the system is cost effective is still not determined.
-
The auditory system is somewhat confusing and cumbersome. It may end up being a good training apparatus but not appropriate for long term use.
-
The electronics package in the below knee segment of the prosthesis presents some problems related to the cosmetic cover which has to allow frequent access for adjustment and battery changes. An attempt is underway at present to replace the instrumented pylon with an instrumented ankle bolt.
-
Investigation is still needed to determine exactly what information is useful. Knee position, for instance, may be more useful information than the direction and magnitude of loading.
*Grace Horton, R.P.T. Department of Surgery, School of Medicine, Department of Medicine, Department of Biomedical Engineering, School of Engineering, Duke University, Durham, N.C.
*David W. Vaughn, C.P.O. Department of Surgery, School of Medicine, Department of Medicine, Department of Biomedical Engineering, School of Engineering, Duke University, Durham, N.C.
*Maret G. Maxwell, Ph.D. Department of Surgery, School of Medicine, Department of Medicine, Department of Biomedical Engineering, School of Engineering, Duke University, Durham, N.C.
*James H. McElhaney, Ph.D Department of Surgery, School of Medicine, Department of Medicine, Department of Biomedical Engineering, School of Engineering, Duke University, Durham, N.C.
*Frank W. Clippinger, M.D. Department of Surgery, School of Medicine, Department of Medicine, Department of Biomedical Engineering, School of Engineering, Duke University, Durham, N.C.
Full Text PDF:
Figure 1:
Figure 2:
Review Status:
Citation
Frank W. Clippinger, M.D. * James H. McElhaney, Ph.D * Maret G. Maxwell, Ph.D. * David W. Vaughn, C.P.O. * Grace Horton, R.P.T. * Linda Bright, R.N. * , “Prosthetic Sensory Feedback Lower Extremity,” Digital Resource Foundation for Orthotics and Prosthetics, accessed November 18, 2024, https://the.oandplibrary.org/items/show/179600.