Orthopedic Walkers: Effect on Plantar Pressures

James A. Birke, P.T., M.S. *
Deborah A. Nawoczenski, P.T., M.Ed. *

Description

https://www.oandplibrary.org/cpo/pdf/1988_02_074.pdf
http://www.oandplibrary.org/cpo/images/1988_02_074/1988_02_074-1.jpg
http://www.oandplibrary.org/cpo/images/1988_02_074/1988_02_074-2.jpg
http://www.oandplibrary.org/cpo/images/1988_02_074/1988_02_074-3.jpg
http://www.oandplibrary.org/cpo/images/1988_02_074/1988_02_074-4.jpg
http://www.oandplibrary.org/cpo/images/1988_02_074/1988_02_074-5.jpg
1988_02_074-6.jpg
1988_02_074-7.jpg
1988_02_074-8.jpg
1988_02_074-9.jpg

Title:

Orthopedic Walkers: Effect on Plantar Pressures

Creator:

James A. Birke, P.T., M.S. *
Deborah A. Nawoczenski, P.T., M.Ed. *

Page Number(s):

74 - 80

Year:

1988

Volume:

12

Issue:

2

Text:

Orthopedic Walkers: Effect on Plantar Pressures

James A. Birke, P.T., M.S. *
Deborah A. Nawoczenski, P.T., M.Ed. *

Introduction

Short leg (SLW) and patellar tendon bearing walkers (PTBW) are orthotic appliances* which have been recently designed as alternative devices to traditional plaster cast immobilization. The indications for use of lower leg walkers include severe ankle sprains, and ankle and foot fractures. Orthopedic walkers are convenient to use, lightweight, and removable to perform joint range of motion or inspect the extremity. Short leg walkers have been shown to be as effective as walking casts in healing stable ankle fractures, and patients treated with short leg walkers have shown significantly less edema, tenderness, and joint stiffness after six weeks of immobilization. The authors feel that orthopedic walkers may also prove to be a beneficial alternative to traditional management of neuropathic fractures and plantar ulcerations, which are commonly seen in diabetes mellitus and Hansen's disease.

Neuropathic foot lesions are the result of abnormal or repetitive stress. Treatment techniques for neuropathic foot conditions should be effective in reducing pressure and shear stress. Traditional methods of treating neuropathic foot lesions include walking casts, fixed ankle braces, and PTB braces. Plaster walking casts and PTB braces have been shown to significantly reduce pressure on the plantar surface of the foot during walking. The total contact walking cast is considered effective in reducing pressure on the foot by redistributing forces on the plantar surface of the foot and lower leg. Several features of PTB orthoses shown to be important in achieving maximal weight bearing reduction on the foot include a rigid closure PTB shell, a heel-shoe clearance of 3/8" to 1", a fixed ankle joint, and a rocker sole. Orthopedic walkers incorporate these same design features to varying degrees which has generated our interest in studying their effectiveness in reducing pressure on the foot.

The SLW has a fixed ankle joint, rocker sole, and a polyurethane liner which is snugly secured to the leg with Velcro® closures (Fig. 1). The PTBW incorporates all the features of the SLW, as well as a non-custom molded, semi-rigid polyethylene PTB shell (Fig. 2).

Figure 1. Short Leg Walker.

Figure 2. Patellar Tendon Bearing Walker.

The effectiveness of the SLW or PTBW in reducing pressure or shear stress on the foot has not previously been studied. The potential value of these devices in managing the neuropathic foot may be evaluated by their effectiveness in reducing pressure and shear stress. Currently, there are unreliable methods for measuring shear stress. However, shear is directly related to the perpendicular forces acting on the foot. Pressure equals the perpendicular forces per unit area. Pressure transducers provide a repeatable measurement of relative pressure inside footwear when the material interfacing with the transducers is controlled.

Purpose

The purpose of this study was to determine the effectiveness of SLW and PTBW in reducing the pressure distribution on the normal foot during walking.

Method

Ten subjects (6 male and 4 female) without a history of foot pathology participated in this study. Capacitive pressure transducers* 2mm thick and 1.5cm in diameter were taped to the first metatarsal head (MTH), third MTH, fifth MTH, and plantar heel of the right foot of each subject (Fig. 3). The foot was covered with a thin cotton stockinette which remained undisturbed during the study. Transducers were calibrated according to the manufacturer's instructions prior to testing each subject. Pressure recordings were made using a four-channel capacitive impedance bridge amplifier* and oscillographic recorder* while subjects walked in a cast shoe (CS-1) (Fig. 4), short leg walker (SLW), patella tendon bearing walker (PTBW), and again in a cast shoe (CS-2). All the walking devices were fabricated by the same manufacturer.* The cast shoe was identical to the foot component of both the SLW and PTBW, utilizing identical rocker out-ersoles and 2.4mm polyurethane material insoles. SLW and PTBW were applied to the leg with a 3/8" heel-shoe clearance. Subjects walked a distance of 100 meters for each treatment condition. The testing order of treatments SLW, PTBW, and CS-2 was randomly assigned to eliminate systematic error.

Figure 3. Pressure transducer placement on selected areas of the foot.

Figure 4. Cast Shoe.

Relative pressure was measured in millimeters of peak to peak chart deflection for 24 steps for each treatment condition. The middle distance of each run was used for analysis in order to eliminate pressure variations due to the acceleration and deceleration phases of each trial. Percent pressure change relative to CS-1 was calculated for treatments SLW, PTBW, and CS-2. Means and standard deviations were computed for treatments at each transducer site. An analysis of variance for repeated measures was used to determine whether treatment differences were significant within each site. Duncan's test was used for post-hoc analysis of means. A significance level of 0.05 was used for comparisons.

Results and Discussion

An analysis of variance (Table I) for mean percent reduction in pressure was highly significant at all sites tested (Fig. 5, Fig. 6, Fig. 7, and Fig. 8). Duncan's test was performed to establish which treatments differed. Significant differences were found between the percent reduction in pressure walking in SLW and PTBW as compared to the CS-2 at all sites. No difference was found between SLW and PTBW at any site. The percent pressure reduction using the walker devices was comparable at all the sites tested.

Table I. Analysis of Variance of Percent Pressure Reduction.

Figure 5. Percent pressure reduction at the first metatarsal head (1 MTH) walking in cast shoe-2 (CS-2), short leg walker (SLW) and patellar tendon bearing walker (PTBW) compared to walking in cast shoe-1.

Figure 6. Percent pressure reduction at the third metatarsal head (3 MTH) walking in cast shoe-2 (CS-2), short leg walker (SLW) and patellar tendon bearing walker (PTBW) compared to walking to in cast shoe-1.

Figure 7. Percent pressure reduction at the fifth metatarsal head (5 MTH) walking in cast shoe-2 (CS-2), short leg walker (SLW) and patellar tendon bearing walker (PTBW) compared to walking in cast shoe-1.

Figure 8. Percent pressure reduction at the heel walking in cast shoe-2 (CS-2), short leg walker (SLW) and patellar tendon bearing walker (PTBW) compared to walking in cast shoe-1.

This study demonstrated the effectiveness of the short leg and patellar tendon bearing walkers as compared to the cast shoe in reducing plantar pressure on the foot. Since all the devices in this study had the same sole design and insole materials, treatment differences must be attributable to proximal orthotic components including the polyurethane liner, fixed ankle uprights, and Velcro® closures. The SLW and PTBW differed only by the polyethylene, non-custom molded patellar tendon cuff. Since no treatment difference was seen between these devices, the PTBW cuff design must not have been effective. However, in follow-up, single subject trials, we were not able to change walking pressures by redesigning the PTBW cuff using polyethylene or plaster custom molded PTB cuffs. An alternative conclusion is that the SLW design alone optimally reduced plantar pressure by the fixed ankle joint and uprights snugly supporting the lower leg and calf.

In this study, orthopedic walkers were equally effective in reducing pressure at all sites tested on the foot. In previous studies, casts were shown to reduce pressure more effectively in the forefoot than the heel, and PTB orthotics reduced pressure more effectively in the heel than the forefoot.

Based on the results of this study, othopedic walkers may be effective devices in the reduction of plantar foot pressure in patients with neuropathic conditions of the foot. There is no evidence to show that the PTBW will be more effective than the SLW. Further study utilizing a patient population is recommended.

Conclusions

Within the scope of this study, it is possible to conclude the following: (1) SLW and PTBW orthopaedic walkers are effective in reducing pressure at the first MTH, third MTH, fifth MTH and heel in normal subjects during walking, and (2) there is no difference in pressure distribution between the SLW and PTBW during walking.

References:

  1. Anderson, J.G., "Treatment and Prevention of Plantar Ulcers," Leprosy Review, 35, 1964, pp. 251-258.
  2. Birke, J.A. and D.S. Sims, "Walking Casts: Effect on Plantar Foot Pressures," Journal of Rehabilitation Research and Developement, 22:3, July, 1985, pp. 18-22.
  3. Brand, P.W., "The Insensitive Foot," Editor M.H. Jahss, Disorders of the Foot, Vol. II, W.B. Saunders, 1982, p. 1266.
  4. Cterctecko, G.C., M. Dhanendran, W.S. Hutton, and L.P. LeQuesne, "Vertical Forces Acting on the Foot of Diabetic Patients with Neuropathic Ulceration," British Journal of Surgery, 68, 1981, pp. 609-614.
  5. Coleman, W.S., P.W. Brand, and J.A. Birke, "The Total Contact Cast: A Therapy for Plantar Ulceration of the Insensitive Foot," Journal of the American Pediatric Medical Association, 74:11, November, 1984, pp. 548-552.
  6. Enna, CD., P.W. Brand, J.K. Reed, and D. Welch, "The Orthotic Care of the Denervated Foot in Hansen's Disease," Orthotics and Prosthetics, 30:1, March, 1976, pp. 33-39.
  7. Gristina, A.G., A.L.W. Thompson, N. Kester, W. Walsh, and J.A. Gristina, "Treatment of Neuropathic Conditions of the Foot and Ankle with a Patellar-Tendon-Bearing Brace," Archives of Physical Medicine and Rehabilitation, 54, December, 1973, pp. 562-564.
  8. Hall, O.C. and P.W. Brand, "The Etiology of the Neuropathic Plantar Ulcer," Journal of the American Pediatric Medical Association, 69:3, March, 1979, pp. 173-177.
  9. Helm, P.A., S.C. Walker, and G. Pullium, "Total Contact Casting in Diabetic Patients with Neuropathic Foot Ulcerations," Archives of Physical Medicine and Rehabilitation, 65, 1984, pp. 691-693.
  10. Lang-Stevenson, A.I., W. Sharrard, R.P. Betts, and T. Duckworth, "Neuropathic Ulcers of the Foot," Journal of Bone and Joint Surgery, (British) 67B, 1985, pp. 438-442.
  11. Lehmann, J.F., CG. Warren, D.R. Pemberton, B.C. Simons, and B.J. DeLateur, "Load-bearing Function of Patellar Tendon Bearing Braces of Various Designs," Archives of Physical and Medical Rehabilitation, 52, August, 1971, pp. 366-370.
  12. Patterson, R.P., and S.V. Fisher, "The Accuracy of Electrical Transducers for the Measurement of Pressure Applied to the Skin," IEEE Transactions on Biomedical Engineering, 26:8, August, 1979, pp. 450-456.
  13. Polakoff, D.R., S.M. Pearce, D.P. Grogan, and W.Z. Burkhead, "The Orthotic Treatment of Stable Ankle Fractures," Orthopedics, 7, 1984, pp. 1712-1715.
  14. Pollard, J.P., and L.P. LeQuesne, "Method of Healing Diabetic Forefoot Ulcers," British Medical Journal, 286, February, 1983, pp. 436-437.
  15. Pollard, J.P., L.P. LeQuesne, and J.W. Tappin, "Forces Under the Foot," Journal of Biomedical Engineering, 5, 1983, pp. 37-40.
  16. Sabato, S., Z. Yosipovitch, A. Simkin, and J. Sheskin, "Plantar Trophic Ulcers in Patients with Leprosy," International Orthopedics, 6, 1982, pp. 203-208.
  17. Soderberg, G., "Follow-up of Application of Plaster-of-Paris Casts for Noninfected Plantar Ulcers in Field Conditions," Leprosy Review, 41, 1970, pp. 184-190.
Footnote 3D Orthopedics, Inc., 10520 Olympic Drive, Dallas, Texas 75220.

Footnote Gulton TR-400a, Gulton Industries, Inc., East Greenwich, Rhode Island.

Footnote Hercules Orthoflex Data System, Allegany Ballistics Lab, Cumberland, Maryland.

Footnote Hercules Orthoflex Data System, Allegany Ballistics Lab, Cumberland, Maryland.


*Deborah A. Nawoczenski, P.T., M.Ed. Deborah A. Nawoczenski, P.T., M.Ed., is Assistant Professor at the Department of Physical Therapy for the College of Allied Health Professions at Temple University, Philadelphia, Pennsylvania 19140.

*James A. Birke, P.T., M.S. James A. Birke, P.T., M.S., is Chief of the Physical Therapy Department at G.W. Long Hansen's Disease Center, Carville, Louisiana 70721.

Full Text PDF:

https://www.oandplibrary.org/cpo/pdf/1988_02_074.pdf

Figure 1:

http://www.oandplibrary.org/cpo/images/1988_02_074/1988_02_074-1.jpg

Figure 2:

http://www.oandplibrary.org/cpo/images/1988_02_074/1988_02_074-2.jpg

Figure 3:

http://www.oandplibrary.org/cpo/images/1988_02_074/1988_02_074-3.jpg

Figure 4:

http://www.oandplibrary.org/cpo/images/1988_02_074/1988_02_074-4.jpg

Figure 5:

TABLE I
http://www.oandplibrary.org/cpo/images/1988_02_074/1988_02_074-5.jpg

Figure 6:

FIGURE 5
http://www.oandplibrary.org/cpo/images/1988_02_074/1988_02_074-6.jpg

Figure 7:

FIGURE 6
http://www.oandplibrary.org/cpo/images/1988_02_074/1988_02_074-7.jpg

Figure 8:

FIGURE 7
http://www.oandplibrary.org/cpo/images/1988_02_074/1988_02_074-8.jpg

Figure 9:

FIGURE 8
http://www.oandplibrary.org/cpo/images/1988_02_074/1988_02_074-9.jpg

Review Status:

Needs Revision- See Trello

Citation

James A. Birke, P.T., M.S. * Deborah A. Nawoczenski, P.T., M.Ed. * , “Orthopedic Walkers: Effect on Plantar Pressures,” Digital Resource Foundation for Orthotics and Prosthetics, accessed December 22, 2024, https://the.oandplibrary.org/items/show/179357.